Overview
High-performance liquid chromatography (HPLC) separates mixtures in solution by pumping a sample at high pressure through a column packed with fine particles. For peptides, reversed-phase HPLC (RP-HPLC) is the standard technique for distinguishing the target compound from by-products of synthesis. Among other things, the resulting chromatogram is used to determine the chromatographic purity of a batch.
Scientific background
Separation principle
All chromatography relies on molecules distributing differently between a stationary phase and a mobile phase. In HPLC, the stationary phase consists of porous particles a few micrometres in diameter, packed tightly into a stainless-steel column. The mobile phase – a solvent mixture – is pumped through at pressures of several hundred bar; UHPLC systems using sub-2 µm particles operate above 1,000 bar. The more strongly a molecule interacts with the stationary phase, the later it leaves the column. The time to the signal maximum is called the retention time.
Reversed-phase HPLC for peptides
In reversed-phase chromatography the stationary phase is non-polar, typically silica bearing octadecyl (C18) or octyl (C8) chains. The mobile phase is water and acetonitrile, frequently with 0.1 % trifluoroacetic acid (TFA) added as an ion-pairing reagent. In gradient mode the organic fraction rises steadily during the run, so that peptides elute in order of their hydrophobicity. Wide-pore materials (around 300 Å) are used for larger peptides so that the molecules can access the pores.
Detection
The peptide bond absorbs mainly in the far-UV region below roughly 230 nm, so detection is usually carried out at 210–220 nm. At 280 nm the aromatic side chains of tryptophan and tyrosine are also recorded. When the column is coupled to a mass spectrometer, the technique is referred to as LC-MS.
Evaluation
The chromatogram plots the detector signal against time. Chromatographic purity is the area of the main peak as a percentage of the sum of all integrated peaks (area per cent). This figure depends on the method: column, gradient, wavelength and integration parameters all influence the result. Substances that co-elute with the main peak or do not absorb UV light – water, salts or counterions, for instance – remain invisible. HPLC purity is therefore not the same as the peptide content of a sample, and a meaningful result always states the method used.
Molecular data
No molecular data is recorded for this entry.
State of research
HPLC has been established for decades and is described in the pharmacopoeias; for peptides, the reversed-phase variant is the reference technique for determining purity. Current developments focus mainly on UHPLC with very small particles, superficially porous (core-shell) materials for faster separations and routine coupling with high-resolution mass spectrometry. Mant et al. (2007) provide a detailed methodological overview of peptide HPLC.
Sources
The sources for this entry are currently being verified by the editorial team and will appear once verified.
Safety and research notes
- Mobile phases frequently contain acetonitrile (highly flammable liquid and vapour, harmful) and trifluoroacetic acid (causes severe skin burns and eye damage). Prepare them only in a fume hood, wearing a lab coat, safety glasses and suitable protective gloves.
- Read the safety data sheets for all solvents and reagents before starting work.
- Collect solvent waste separately in line with the laboratory’s waste rules and dispose of it according to local regulations.
- The methodology described here serves solely the analytical characterisation of research material.