Overview
LC-MS couples a liquid-chromatographic separation directly to a mass spectrometer. Each component eluting from the column is ionised and recorded according to its mass-to-charge ratio. For peptides, a single run can confirm the identity of the main component and assign by-products by their mass.
Scientific background
Set-up
An LC-MS system consists of an HPLC or UHPLC unit, an ion source and a mass analyser. Peptides are almost always ionised by electrospray ionisation (ESI): the eluate is sprayed under high voltage, the solvent evaporates and the molecules pass into the gas phase as charged ions. Peptides typically pick up several protons and appear as a series of multiply charged ions [M+nH]ⁿ⁺. The mass of the neutral molecule can be calculated from this charge series (deconvolution).
Mass analysers
Quadrupole, ion-trap, time-of-flight (TOF) and Orbitrap analysers are all widely used. They differ mainly in resolution and mass accuracy. High-resolution instruments determine masses to within a few ppm and cleanly separate the isotope signals of a peptide. In tandem mass spectrometers (MS/MS), selected ions are additionally fragmented; the resulting b and y fragments allow conclusions to be drawn about the amino acid sequence.
Use in peptide analysis
Identity testing compares the measured mass with the mass calculated from the sequence. By-products of synthesis often reveal themselves through characteristic mass differences: a deletion sequence lacks the mass of one amino acid residue, oxidation (of methionine, for example) adds 16 Da, deamidation roughly 1 Da, and a tert-butyl protecting group that has not been removed 56 Da. Because such by-products often elute close to the main peak, the combination of separation and mass determination is particularly valuable for purity analysis: a signal that appears only as a shoulder in the UV chromatogram can often be assigned unambiguously by its mass.
Limitations
Signal intensity in a mass spectrum is not a direct measure of quantity, because different substances ionise with different efficiency. Quantification therefore usually relies on the HPLC UV signal or on calibration with reference material. Isomers of identical mass – peptides containing a D-amino acid or an isoaspartate residue, for instance – cannot be distinguished by the mass spectrometer alone; chromatographic separation is what matters here. The mobile phase is also a compromise: TFA improves peak shape but suppresses ionisation, which is why formic acid is frequently used for LC-MS.
Molecular data
No molecular data is recorded for this entry.
State of research
LC-MS is routine in peptide and protein analysis. Its foundation was the development of electrospray ionisation for large biomolecules in the late 1980s (Fenn et al., 1989). Current developments include high-resolution instruments in routine use, ion mobility spectrometry as an additional separation dimension and automated data-evaluation software.
Sources
The sources for this entry are currently being verified by the editorial team and will appear once verified.
Safety and research notes
- The ESI source operates at high voltage; carry out maintenance only with the instrument switched off and in accordance with the manufacturer’s instructions.
- Nitrogen used as nebulising and drying gas can displace oxygen in poorly ventilated rooms; check the gas supply and room ventilation regularly.
- The notes on HPLC solvents (acetonitrile, acid additives) apply to the mobile phases: consult the safety data sheets and collect waste separately.
- Handle samples exclusively as research material and label them clearly.